    The LIGHTMED LIGHTLas 532 green laser photocoagulator can be used in both traditional continuous wave or SP-Mode (sub-threshold technology) for optimal patient outcomes. Additionally, LIGHTMED’s unique modular design allows for the integration of SLT and/or YAG functionalities at any time.

    Also available with integrated slit lamp, table, and smart cart.

    FEATURES:

    • Conventional CW and Sub-Threshold SP-MODE (micro pulse) Laser Mode (adjustable platform)
    • Laser Cavity Bonding
    • Instant Duty-Cycle Circularity
    • Superior Laser Crystal Coating
    • Continuous System Monitoring
    • Wireless foot pedal with foot Power Control
    • 7″ LCD Touch Screen Interface
    • Safety Glasses
    • Packed into a carrying case

    ACCESSORIES & DELIVERY OPTIONS

    • Laser Indirect Ophthalmoscope
    • Endoprobes
      • Sizes 20G, 23G, 25G, 27G
      • Types: Straight, Illuminated, Flexible
    • Slit Lamp Delivery Attachment with Haag Streit Slit Lamp
    • Fixed Safety Filter for Leica 841/501 SM
    • Fixed Safety Filter for Zeiss / Moeller SM

    UPGRADE OPTIONS

      SPECIFICATIONS:

      • Laser system: Diode Pump Solid State (DPSS) true CW and SP-Mode™
      • Safety classification: Class 4
      • Wavelength: 532 nm (green)
      • Power output: 50 mW – 2.0 W and 50 mW – 4.0 W (ENT optional)
      • Max power at cornea: 2.0 W (Endo, LIO, and SLA at all spot sizes)
      • Pulse duration: 0.01 – 3.0 seconds
      • Pulse interval: 0.01 – 3.0 seconds, and continuous
      • Sub-threshold laser therapy (SP-Mode) settings:
        • Duration: 150 μs – 600 μs (in 50 μs increments)
        • Duty cycle: 5% – 30% (in 2.5% increments)
        • Period: 1400 μs – 1850 μs (in 50 μs increments)
      • Aiming beam: Laser diode 653 nm (red), 0.1 – <1 mW
      • Safety classification: Class 2
      • Dimensions: (laser console) 13 cm (H) x 36 cm (W) x 33 cm (D), 5.1 in (H) x 14.5 in (W) x 12.9 in (D)
      • Weight (laser console): 12 kg, 26.4 lbs.
      • Power requirements: 100 – 340 VAC, 50/60 Hz auto-ranging